Precision control of oxygen and carbon dioxide for micro-scale cell culture
The 颇尔 Micro-24 Gas Micro Mixer accessory creates additional gas regime control capability at the microbioreactor scale, increasing the precision of control of CO2 exposure for sensitive mammalian cell culture applications.A very simple touch screen interface controls the mixing of up to three control gases - Oxygen, Carbon Dioxide, and Nitrogen/ Normal Air - with precision low-volume thermal mass flow controllers (Figure 2 on Performance tab).The user can access pre-set standard configurations, or quickly and easily create and store custom mixes of 2 or 3 gases.The device supports very precise pH and dissolved oxygen control while maintaining optimal CO2 concentrations in culture, stripping excess CO2 and maintaining minimal CO2 (Figure 3 on Performance tab).
特色及优点
特点 | 优势 |
Easily mixes standard laboratory gases (O2, CO2, N2 or Air) | Eliminates acquisition and handling of mixed gas cylinders, simplifies operations |
Allows optimization of gas control strategies | Reduces evaporation, process instability, and culture variability |
More precise control of monitored (pH) and unmonitored (CO2) process parameters |
Prevents in process CO2excursions
|
Touch screen control interface | Easily create, edit, and store gas mixes, rapidly access “standard” and saved gas mixes |
Compact design | Laboratory bench, either behind 颇尔 Micro-24 Microreactor or beside 颇尔 Micro-24 Microreactor. (Stand supplied to add height for use behind 颇尔 Micro-24 Microreactor) |
LED strip under lid | Illuminating strip under the lid showing operating state which can be seen from distance |
Automatic purge | Performs a purge on all the gas lines for a set period of time before use |
Status messages display on screen | Status messages are displayed on screen during operation to update the user. (Refer to manual for descriptions) |
Gas Line Schematic Diagram


电源 | 110-240 VAC, 50/60 Hz self-switching |
Input Gas Supply |
Analytical Grade Oxygen, CO2, and N2 /Air Max input pressure 8.27 bar/120 psi Connections:6mm Push-to-Connect 注意事项:The inlet pressure must be at least 750 mbar (10.9 psi) higher than the required output pressure |
Mixed Gas Outputs |
电话:0.25 – 25 sccm Purge/ Inert:0.1-10 sccm CO2 and 0.25-25 sccm N2 / Air DO:0.1-10 sccm CO2 + O2 and 0.25-25 sccm N2 /Air Connections:6 mm Push-to-Connect |
Gas Flow Accuracy and Stability |
Accuracy set points Set point 0 – 10% Accuracy +/- 1.0% Set point 10 – 50% Accuracy +/- 1.0% Set point 50 – 100% Accuracy +/- 1.0% Stability once gas mix has been set Set point 0 – 10% Stability +/- 0.5% Set point 10 – 50% Stability +/- 0.5% Set point 50 – 100% Stability +/- 0.5% Stability achieved within 30 seconds |
Environmental Requirements |
Operating Range +5°C to +40°C Altitudes up to 2,000 m (6,500 ft) Maximum relative humidity of 80% for temperatures up to +31°C, decreasing to 50% at +40°C, with no condensation |
合规性
EMC Directive | 2004/108/EC |
Low Voltage Directive | 2006/95/EC |
Waste Electrical and Electronic Equipment | 2002/96/EC |
Reduction of Hazardous Substances | 2002/95/EC |
Safety requirements for electrical equipment for measurement, control and laboratory use - Part 1:General Requirements | BS EN 61010-1:2001 |
Electrical Equipment For Measurement, Control and Laboratory Use - EMC Requirements - Part 1:General Requirements | BS EN 61326-1:2006 |
Specification for Degrees of Protection provided by Enclosures (IP Code) | BS EN 60529:2000 |
图 2.Touch Screen Control Interface

Touch Screen set to provide 3 separate gas mixes:pH control mix with 50% N2 and 50% CO2, Purge mix of 5% CO2 and 95% N2, and Dissolved Oxygen Control mix with 21% O2, 5% CO2, and 74% N2
图 3.Gas Micro Mixer maintains optimal CO2 during culture

Gas mixtures maintain optimal Carbon Dioxide levels with minimal impact to feedback control of Dissolved Oxygen and pH during culture.
描述 | Part Number |
颇尔 Micro-24 气体微混合器 | MRT-GMM-1 |