pall,
颇尔,Fluorodyne II 膜的 AcroPak 20 过滤器和 AcroPak 200
囊式过滤器
Efficient Processing of Liquid
Volumes Typically Up To 20 L
-
Fluorodyne II membrane is compatible with aqueous and many organic solvents
-
AcroPak 20 filters are designed to add convenience to small- and medium-volume filtrations
-
AcroPak 200 Fluorodyne II capsules provide fast processing of clean batches
-
Double layer sterilizing membrane enhances reliability
-
上游排气可防止蒸气凝结
-
Manufactured without the use of adhesives to minimize extractables
-
Pre-sterilized by gamma irradiation

-
Sterile filtration of media and buffers
-
Point-of-use filtration for lab water
-
Suitable for viscous, particulate solutions and serum containing media
-
Applicable for many fluids containing dilute proteins or preservatives
-
Scalable
材料成分
-
Filter Media:Fluorodyne II membrane (hydrophilic PVDF)
-
Housing, Vent Plug:聚丙烯
-
Sealing Technology:Thermal bonding
-
Filling Bell:聚碳酸酯
孔径
有效过滤面积
Dimensions (Nominal)
-
AcroPak 20 过滤器
Housing Length:8.3 cm (3.3 in.)
Housing Diameter:6.7 cm (2.7 in.)
-
AcroPak 200 囊式过滤器
Overall Length:10.5 cm (4.1 in.)
Housing Diameter Without Vent:5.3 cm (2.1 in.)
Housing Diameter With Vent:6.7 cm (2.6 in.)
Inlet/Outlet Connections
-
6.4 - 12.7 mm (1/4 - 1/2 in.) diameter stepped hose barb with female luer slip interior and filling bell outlet
Maximum Operating Temperature and Pressure
-
60 °C (140 °F) at 1.0 bar (100 kPa, 15 psi)
-
4.1 bar (410 kPa, 60 psi) at 21 - 24 °C (70 - 75 °F)
Recommended Integrity Test Minimum Bubble Point – Water
-
0.2 μm Fluorodyne II membrane:3.2 bar (320 kPa, 46 psi)
60% IPA/40% H2O (v:v)
-
0.1 μm Fluorodyne II membrane:1.8 bar (180 kPa, 26 psi)
Typical Water Flow Rate mL/min at 0.1 bar (10 kPa) [mL/min/psi]
-
AcroPak 20 过滤器
0.1 μm Fluorodyne II Membrane:13 (9)
0.2 μm Fluorodyne II Membrane:26 (18)
-
AcroPak 200 囊式过滤器
0.2 μm Fluorodyne II Membrane:240 (170)
Typical Hold-Up Volume (with 10 psi air purge)
-
AcroPak 20 Filters < 2.5 mL
-
AcroPak 200 Capsule < 6 mL
Bacterial Retention
-
Lot samples retain a minimum of 107 cfu/cm2 of B. diminuta per modified ASTM F838, current revision
Endotoxin Level
-
< 0.25 EU/mL using Limulus Amoebocyte Lysate (LAL) test
Biological Safety
-
Passes United States Pharmacopeia (USP) Biological Reactivity Test, In Vivo <88>
Sterilization
-
Sterilized by gamma irradiation and individually packaged
-
AcroPak 20 filters can withstand one autoclave cycle at 131 °C (268 ºF) for 30 min