For Buffers, Tissue Culture Media,
and Other Biological Fluids
-
Newer technology recommended for simplified scale-up
-
High filtration area and compact size are ideal for upscale trials
-
Minimal hold-up volume
-
Good wettability for reliable integrity test
-
Sealed membrane and housing uses fusion technology to eliminate potential extractables
-
Gamma sterilized, non-pyrogenic, and provided with a removable filling bell (except sanitary flange option)
-
Hydrophilic polyethersulfone membrane for low adsorption and wide chemical compatibility
-
Very high flow rates and consistently higher total solution throughputs due to superior porosity over other membranes

-
Designed to quickly process difficultto- filter solutions within 5-21 L, such as serum, serum-supplemented culture media, and ascites fluid
-
Ideal in situations where rapid filtration or short processing times are essential
材料成分
-
Filter Media:Supor EKV [hydrophilic polyethersulfone (PES)] membrane
-
Housing, Vent Plug, and Support Material:聚丙烯
-
Filling Bell:Polycarbonate (no filling bell on sanitary option)
-
Sealing Technology:Thermal bonding without adhesives
孔径
有效过滤面积
尺寸
-
Housing Length:12094:10.5 cm (4.1 in.);12095:7.3 cm (2.9 in.)
-
Housing Diameter:5.3 cm (2.1 in.) without vent, 6.7 cm (2.6 in.) with vent
Inlet/Outlet Connections
-
12094:6.4 - 12.7 mm (1/4 - 1/2 in.) stepped hose barb inlet and outlet
-
12095:13 mm (1/2 in.) sanitary flange inlet/outlet
Typical Hold-Up Volume (with 10 psi air purge)
Maximum Operating Temperature
-
60 °C (140 °F) at 2.1 bar (210 kPa, 30 psi)
最大操作压力
-
4.1 bar (410 kPa, 60 psi) at ambient temperature
Recommended Integrity Test Minimum Bubble Point - Water
-
> 3.32 bar (332 kPa, 48 psi)
Typical Liquid Flow Rate
-
350 mL/min/0.1 bar (241 mL/min/psi)
Bacterial Retention
-
Lot samples retain a minimum of 107 cfu/cm2 of B. diminuta per modified ASTM F838, current revision
Endotoxin Level
-
< 0.25 EU/mL using Limulus Amoebocyte Lysate (LAL) test
Biological Safety
-
Passes United States Pharmacopeia (USP) Biological Reactivity Test, In Vivo <88>
Sterilization
-
Sterilized by gamma irradiation and individually packaged.If desired, autoclave once at 131 ºC (268 ºF) for 30 minutes.Water wet capsule prior to autoclaving.