Membrane Filters for Proven Applications
- Obtain accurate fills, uniform flow, and good throughput.Built-in air vent allows elimination of trapped air bubbles
- Integrity testable to ensure product and system integrity.
- Sterilization by gamma irradiation eliminates potential contamination by EtO residuals.
- Use for filtration of batches ranging from 1 to 10 liters (Micro Culture Capsule) and 10 to 20 liters (Culture, Maxi Culture, and Maxi Capsules) for the sterilization of buffers and cell culture media
材料成分
- Filter Media:HT Tuffryn membrane (polysulfone)
- Housing and Filling Bell:聚碳酸酯
- Vent Plug and Core:聚丙烯
- Filter Support:聚酯
- Sealing Materials:Polyurethane
有效过滤面积
- PN 12158:300 cm2
- PN 12140 and 12170:500 cm2
- PN 12141:1290 cm2
Inlet/Outlet Connections
- PN 12140, 12170, and 12141:Hose barb fittings accept 12.7 mm (1/2 in.)ID tubing
- PN 12158:Stepped hose barbs 6.4 - 9.5 mm (1/4 - 3/8 in.) diameter
Typical Hold-Up Volume with 0.7 bar (70 kPa, 10 psi) air purge
- PN 12158:15 毫升
- PN 12140 and 12170:25 毫升
- PN 12141:113 毫升
Maximum Operating Temperature
- 100 °C (212 °F)
最大操作压力
- 3.4 bar (340 kPa, 50 psi) at ambient temperature
Recommended Integrity Test Minimum Bubble Point – Water
- 2.4 bar (240 kPa, 35 psi)
Typical Water Flow Rate L/min at 0.7 bar (70 kPa, 10 psi)
- PN 12158:1.2
- PN 12140, and 12170:2.0
- PN 12141:5.2
Bacterial Retention
- Lot samples retain a minimum of 107 cfu/cm2 of B. diminuta per modified ASTM F838, current revision
Endotoxin Level
- < 0.5 EU/mL using Limulus Amoebocyte Lysate (LAL) test
Biological Safety
- Passes United States Pharmacopeia (USP) Biological Reactivity Test, In Vivo <88>
Sterilization
- Sterilized by gamma irradiation
- If desired, autoclave once only prior to use at 121 - 123 °C (250 - 253 °F) for a maximum of 20 min