Robust Chemical Resistance Provides
Compatibility with Harsh Organics and Aqueous Solutions
-
Chemically resistant materials provide a stable platform to process samples in organic solvents
-
Inert materials of construction ensure complete recovery of samples and low non-specific binding
-
New well design results in faster, more uniform filtration rates across the plate and reduces hold-up volume
-
Innovative outlet tip geometry provides direct flow of samples into receiver plate without concerns of cross-contamination
-
新陈代谢研究
-
Molecular or drug synthesis reactions
-
Sample prep using solvents
-
Aggressive filtration applications
材料成分
-
Filter Media:PTFE (polytetrafluoroethylene) and GHP(hydrophilic polypropylene) membrane
-
Plate Housing:Polyproylene
-
Lid:Polystyrene
尺寸
-
Length:12.8 cm (5.0 in.)
-
Width:8.6 cm (3.4 in.)
-
Height With Lid:1.8 cm (0.7 in.) (350 μL only)
-
Height Without Lid:350 μL, 1.4 cm (0.6 in.);1 mL:3.3 cm (1.3 in.);2 mL:4.7 cm (1.9 in.)
Well-Bottom Area
Recommended Working Volume
-
350 μL: < 300 μL
-
1 mL: < 900 μL
-
2 mL: < 1.9 mL
Recommended Operating Vacuum
Recommended Centrifugal Force
Plates Maintain Solvents for Extended Incubations
AcroPrep Advance 0.2 and 0.45 µm PTFE Membrane Filter Plates
|
|
|
Ethanol, 100% |
R |
R |
R |
|
Methanol,100% |
R |
R |
R |
|
ACN, 100% |
R |
R |
R |
|
DMSO, 100% |
R |
R |
R |
|
Hexane, 100% |
R |
R |
E |
|
|
This solvent retention table shows results reported for 200 μL of liquid with 30 min., 2 hr. and 24 hr. incubations at room temperature in a humid chamber.
R = fully retained
E = fully retained, completely evaporated before 24 hr. mark
n = 24 (repeated 3 times).
AcroPrep Advance 0.2 and 0.45 µm GHP Membrane Filter Plates
|
|
|
Ethanol, 100% |
R |
R |
E |
|
Methanol,100% |
R |
R |
E |
|
ACN, 100% |
R |
R |
E |
|
DMSO, 100% |
R |
R |
R |
|
Hexane, 100% |
R |
E |
E |
|
|
This solvent retention table shows results reported for 200 μL/well (350 μL plates) and 300 μL/well (1 mL plates) of liquid with 30 min., 2 hr. and 24 hr. incubations at room temperature without a humid chamber.
R = fully retained
E = fully retained, completely evaporated before 24 hr. mark
n = 24 (wells/solution).
注意事项:To help prevent evaporation during long incubation periods, the use of a humid chamber is recommended.