Supor® EBV polyethersulfone membrane cartridges are validated, 0.2 µm sterilizing-grade filters with a unique combination of
颇尔 Ultipleat® filter construction and built-in prefiltration to give longer filter life and lower filtration costs.该过滤器适合各种流体的除菌过滤,包括缓冲液、生物学流体、组织培养基、眼用药品产品和其它多种产品。The low protein and preservative binding of the Supor polyethersulfone membranes also ensures maximum transmission of active ingredients.
The Mini Kleenpak™ KA02 capsules are compact filters used in the laboratory for volumes of 2 L to 50 L in process development, and in pilot and manufacturing scale operations if processing requirements do not demand a large filter.These filters are the smallest capsule filters in the UpScaleSM Program range to utilize a pleated membrane, and are excellent for modelling filter performance at large production scale.
Key Features and Benefits
-
Encapsulated format for higher flexibility, minimized cleaning and low installation costs
-
Minimal hold up volume
-
Ideal for upscale trials
-
Hydrophilic polyethersulfone membrane for low adsorption and wide chemical compatibility
-
Built-in prefilter layer for long-life and low filtration costs
-
High-strength design allows for multiple autoclave cycles and extended use
Quality Standards
-
Manufactured for use in conformance with cGMP
-
100% integrity tested
-
ISO 9000 Certified Quality System
-
Meets USP Biological Reactivity Test, in vivo, for Class VI-121 °C Plastics
-
Every filter tested during manufacture.Test correlated to microbial retention
-
Certificate of Test provided includes:
-
Fabrication Integrity
-
Bacterial Retention
-
Materials of constructions
-
Effluent quality for cleanliness, TOC and Water Conductivity, pH and Pyrogens
材料成分
|
Membrane Layers |
Dual layer polyethersulfone (PES) |
Support and Drainage |
聚丙烯 |
Capsule |
聚丙烯 |
排气 |
聚丙烯 |
Sealing Technology |
Thermal bonding without adhesive |
|
Operating Parameters1
|
最大操作压力 |
4.1 bar (60 psi) at 30oC |
Maximum Operating Temperature |
60oC at 2.1 bar (30 psi) |
|
1 In compatible fluids which do not soften, swell or adversely affect the filter or its materials of construction
Sterilization
|
Autoclaving at 125oC |
3 cycles x 30 minutes |
Gamma irradiation |
Maximum 50 KGy* |
|
* For G option only
Warning:Pre-sterilized products should not be resterilized.This product must not be sterilized in-situ by passing steam through under pressure
Extractables
Nominal Dimensions
|
Height |
10.4 cm (4.1 in.) |
Diameter |
5.1 cm (2.0 in) |
|
Typical Effective Filtration Area
Typical Water Flow
|
KA02EBVP2G |
Suitable for autoclaving or Gamma irradiation with stepped hosebarb ½ in. to ¼ in. connections and filling bell*, 3 per box |
KA02EBVP2S |
Pre-sterilized hosebarb ½ in. to ¼ in. connections and filling bell*, 3 per box |
|